KANE 3100 Series Differential Pressure Meter |
|
|
|
Page 3 of 3
Technical Specification
KANE3100-1 Specifications
KANE3100-1 includes timed let-by, stabilisation and tightness test with log and print facility. Duration of the let-by, stabilisation and tightness test can be set by the user.
|
Nominal Pressure Range
|
Measured Range
|
Resolution
|
Accuracy
|
|
± 80 mBar
Max. overange without damage to the sensor is 5 times nominal range
|
± 0.2 mBar
± 1 mBar
± 80mBar
|
0.001 mBar
0.001 mBar
0.01 mBar
|
±0.005 mBar
±0.03 mBar
± 3% of reading
|
KANE3100-2 Specifications
KANE3100-2 includes timed let-by, stabilisation and tightness test with log and print facility. Duration of the let-by, stabilisation and tightness test can be set by the user.
|
Nominal Pressure Range
|
Measured Range
|
Resolution
|
Accuracy
|
|
± 160 mBar
Max. overange without damage to the sensor is 5 times nominal range
|
± 0.2 mBar
± 19.999 mBar
± 160 mBar
|
0.003 mBar
0.003 mBar
0.01 mBar
|
±0.02 mBar
±0.05 mBar
± 3% of reading
|
KANE3100-5 Specifications
KANE3100-5 includes manual log and auto log facilities. Live and stored tests can be printed.
|
Nominal Pressure Range
|
Measured Range
|
Resolution
|
Accuracy
|
|
± 400 mBar
Max. overange without damage to the sensor is 5 times nominal range
|
± 300 mBar
± 400 mBar
|
0.01 mBar
0.1 mBar
|
± 3% of reading
|
|
Parameter
|
Description
|
|
Dimensions: Weight
|
285 grams with battery
|
|
Handset
|
150mm x 72mm x 23mm (172mm including spigots)
|
|
Ambient Operating Range
|
+0ºC to +45ºC / 10% to 90% RH non-condensing
|
|
Power Supply
|
9 Volt PP3 alkaline battery
|
|