Page 2 of 2
Technical Specification
|
Parameter
|
Range
|
Resolution
|
Accuracy
|
| Temperature Measurement |
|
Flue Temperature
|
0-600°C
|
0.1°C
|
±2.0°C ±0.3% reading
|
|
Inlet Temperature (Internal sensor)
|
0-50°C
|
0.1°C
|
±1.0°C ±0.3% reading
|
|
Inlet Temperature (External sensor)
|
0-600°C
|
0.1°C
|
±2.0°C ±0.3% reading
|
| Gas Measurement |
| Oxygen |
0-21% |
0.1% |
±0.2%*1 |
| Carbon Monoxide |
0-2,000ppm nom
4,000ppm max for 15 mins
|
1ppm |
±10ppm <100ppm*1 ±5% reading
|
|
Carbon Dioxide*2
|
0-30% |
0.1% |
±0.3% reading |
| Efficiency*2 |
0-99.9% |
0.1% |
±1.0% reading |
| Excess Air*2 |
0-250% |
0.1% |
±0.2% reading |
| CO/CO2 ratio *2 |
0-0.999
|
0.0001 |
±5% reading
|
| Pressure (differential) |
|
Nominal range ± 80 mBar
Maximum over range without damage to sensor is ± 400 mBar
|
± 0.2 mBar
± 1 mBar
± 80 mBar
|
0.001 mBar
0.001 mBar
0.01 mBar
|
±0.005 mBar
±0.03 mBar
± 3% of reading
|
|
Storage Capacity:
|
99 Combustion Tests
20 Pressure Tests
20 Let-by / Tightness Tests
20 Temperature Tests
20 Room CO Tests
|
|
Pre-programmed Fuels
|
Natural gas, Propane, Butane, LPG, Light Oils (28/35 sec)
|
| Dimensions |
|
Weight:
Handset:
Probe:
|
0.8kg (1.8lb) handset with boot
200 x 45 x 90mm (7.9" x 1.8" x 3.5")
L300mm (11.8") x dia 6mm (0.25") with 240mm (9.4") long stainless steel shaft, type K thermocouple and 3m (9’) long neoprene hose
|
|
Ambient Operating Range
|
0°C to +40°C 10% to 90% RH non-condensing
|
|
Battery Type / Life
|
4 AA cells
>12 hours using Alkaline AA cells
|
|
Chargers (optional)
|
220v fast charger, for NiMH batteries only
12v in vehicle charger, for NiMH batteries only
|
*1 Using dry gases at STP
*2 Calculated
|