How do I change my KANE3500 let by, stabilisation & tightness intervals?
How do I change my KANE3500 let by, stabilisation & tightness intervals?
Your KANE3500 default interval times are:
1 minute Let by – TIME 01
1 minute Stabilisation – TIME 02
2 minute Tightness – TIME 03
You can change these intervals to any time up to 99 minutes as follows:
Press ON/OFF to turn on your KANE3500
Press MENU then Up / Down arrow buttons to select PRESSURE & press Enter / Arrow button
Press Up / Down to select REPORTS & press Enter / Arrow
Press Up / Down to select SETTINGS & press Enter / Arrow key
The display shows the Let by time:
TIME 01
TIME 00
Use Up / Down to select your time then Enter to move to the next digit & repeat - The cursor indicates which digit will change. Press Enter again to move to TIME 02 & 03
Repeat the above steps to change Stabilisation (TIME 02) & Tightness (TIME 03)
These times are stored in memory until you change them or remove the battery.